The Raman spectroscopy for MEA-CO 2-H 2 O system, transferring the measurement data into Matlab/Labview, converting data into concentration values and presenting the results in a graphical user interface. This software based platform makes the Raman spectroscopy to be accessed as a real-time instrument in CO 2 capture plants. Keywords: CO. The TSL system is a Kaiser Raman RXN1 spectrometer.22 The Kaiser uses a 785 nm NIR gallium arsenide (GaAs) diode as the laser source, with a variable power output of 10–400 mW. Spectral coverage is from 100 to 3450 cm 1 Raman shift, with resolution of 5 cm 1. The Analytical Sample Compart-ment (HLSC) used to collect the spectra incorporates. Major progress toward widespread Raman use, 1985-2005:. 10. Sensitivity increase, 100-500 X in SNR. Compact, low power, integrated systems available.
Kaiser Raman Spectroscopy Rxn1 User Manual Online
Kaiser Raman Spectroscopy Rxn1 User Manual Pdf
Specifications: | Laser Wavelength: | 532nm, 785nm, 1000nm | Spectral Coverage: | 150-4375 cm–1 (532nm), 150-3425 cm–1 (785nm), 200-2400 cm–1(1000nm) | Operating Temperature: | +5 to +30°C | Storage Temperature: | –15 to +65°C | Relative Humidity: | 20–80%, Non-condensing | Warm-up Time: | 30 Minutes |
Electrical Data - Base Unit: | Input Voltage: | 110-240 VAC, 50-60Hz Standard | Max Power: | <400 Watts Max (startup), <150 Watts Typical | User Interface: | Approved PC Operating Windows 10 |
| Physical - Base Unit: | Mounting: | Rack Mounted | Dimensions (width × depth × height): | 19' × 24' × 10.5' (48 cm × 61 cm × 26.7 cm) | Weight (base unit): | 61 lbs. (28 kg) |
Sample Interface: | Solids: | • PhAT Probe Head for Hybrid analyzers only (Non-contact or Immersion Optic) | Liquids: | • Pilot™-E (Process) • WetHead™ (Process / Process Development) • RamanRxn™ Probe (Immersion Optic for R&D and Development) | Gas: | • AirHead™ | Extrusion: | • WetHead™ (Process / Process Development) • RamanRxn™ Probe (Extruder Optic for R&D and Development) |
|